Effects of Light Curing Modes and Resin Composites on Temperature Rise under Human Dentin: An in vitro Study

Küçük Resim Yok

Tarih

2008

Yazarlar

Hubbezoglu, Ihsan
Dogan, Arife
Dogan, Orhan Murat
Bolayir, Giray
Bek, Bulent

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

JAPANESE SOC DENTAL MATERIALS DEVICES

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

The influence of three curing modes of a high-powered LED curing unit on temperature rise under 2-mm-thick dentin was investigated during the polymerization of resin composite samples of Admira, Filtek P60. Premise, Tetric Flow, Tetric Ceram, and Filtek Z250. Ninety standard specimens were prepared. The bonding agents and resin composites were cured with standard, pulse, or soft-start mode (n=5 for each curing mode). Temperature rise was measured using a type 1, thermocouple. Data were analyzed by two-way ANOVA and Tukey's test. Soft-start curing led to statistically higher temperature rises compared than the other two modes. The highest temperature rise was observed for Admira and Tetric Flow cured with soft-start mode. The lowest temperature rise was observed for Premise cured with pulse mode. However, temperature rise did not reach the critical value that can cause pulpal damage by virtue of a prominent safety feature of the high-powered LED LCU, which ensures that no excessive heat is produced by all the three curing modes.

Açıklama

Anahtar Kelimeler

Temperature rise, Composites, Polymerization

Kaynak

DENTAL MATERIALS JOURNAL

WoS Q Değeri

Q4

Scopus Q Değeri

Q1

Cilt

27

Sayı

4

Künye