Evaluation of the compatibility of Phoenix 100 and Microflex LT MALDI-TOF MS systems in the identification of routinely isolated microorganisms in the clinic microbiology laboratory
Tarih
2016Yazar
Celik, CemUysal, Elif Bilge
Tutar, Ugur
Erturk, Rahsan
Bakici, Mustafa Zahir
Gozel, Mustafa Gokhan
Üst veri
Tüm öğe kaydını gösterÖzet
Background: Matrix-assisted laser desorption/ionization time-of flight mass spectrometry (MALDI-TOF MS) is a quick, reliable, and efficient system for identifying microorganisms. Many centers that use the Phoenix 100 system today may adopt aMALDI-TOF MS system in the future. Our laboratory recently undertook this pivot. The present study evaluates the reproducibility of species identifications made by the Phoenix 100 and MALDI-TOF MS systems, during a period of transitioning laboratory instrumentation. Methods: Eight hundred and twelve microbial isolates, from aerobic cultures of different clinical samples, were identified simultaneously with Phoenix 100 (Becton Dick-inson, Sparks, MD, USA) and a Microflex LT MALDI-TOF MS (Bruker Daltonics, Bremen, Germany) devices. Results: Both systems made identical species assignments for 98.9%, 92.1%, 95.1%, and 93.1% of Gram-negative isolates, catalase-positive Gram-positive cocci isolates, cat-alase-negative Gram-positive cocci isolates, and Candida isolates, respectively. Conclusions: Identifications made by two instruments commonly used in microbiology laboratories, the Phoenix 100 and the Microflex LT MALDI-TOF MS, are highly consistent.