dc.description.abstractIn this study, the optimal conditions for the electrodeposition of nickel foam-like morphology over a pencil graphite electrode for glucose sensor application are determined. The effects of NiSO4 and NH4Cl concentrations, current density, and electrodeposition time were investigated in detail. The electrochemical performance of the nickel deposit for sensing glucose is investigated by cyclic voltammetry and chronoamperometry in alkaline solutions. Under optimized conditions, the sensitivity of the sensor toward glucose oxidation was 125.9 μA/μM.cm2 and showed a linear response in a wide range of glucose concentrations, from 2 to 8000 µM. We conclude that the nickel foam-like electrode has the potential to be used as a glucose sensor due to its superior features such as low cost and fast deposition time.tr
dc.description.abstractIn this study, the optimal conditions for the electrodeposition of nickel foam-like morphology over a pencil graphite electrode for glucose sensor application are determined. The effects of NiSO4 and NH4Cl concentrations, current density, and electrodeposition time were investigated in detail. The electrochemical performance of the nickel deposit for sensing glucose is investigated by cyclic voltammetry and chronoamperometry in alkaline solutions. Under optimized conditions, the sensitivity of the sensor toward glucose oxidation was 125.9 μA/μM.cm2 and showed a linear response in a wide range of glucose concentrations, from 2 to 8000 µM. We conclude that the nickel foam-like electrode has the potential to be used as a glucose sensor due to its superior features such as low cost and fast deposition time.tr
dc.publisherSpringertr
dc.rightsinfo:eu-repo/semantics/closedAccesstr
dc.titleEnzyme-free nickel electrochemical glucose sensor fabricated on pencil graphite electrode by electrodepositiontr
dc.typearticletr
dc.contributor.departmentMühendislik Fakültesitr
dc.contributor.authorID0000-0003-4026-867Xtr
dc.identifier.volume34tr
dc.identifier.issue645tr
dc.identifier.endpage13tr
dc.identifier.startpage1tr
dc.relation.publicationcategoryUluslararası Hakemli Dergide Makale - Kurum Öğretim Elemanıtr