Temperature Rises During Application of Er:YAG Laser Under Different Primary Dentin Thicknesses

Küçük Resim Yok

Tarih

2013

Yazarlar

Hubbezoglu, Ihsan
Unal, Murat
Zan, Recai
Hurmuzlu, Feridun

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

MARY ANN LIEBERT INC

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

Objective: The present study investigated the effects of the Er:YAG laser's different pulse repetition rates on temperature rise under various primary dentin thicknesses. Background data: The Er:YAG laser can be used for restorative approaches in clinics and is used to treat dental caries. There are some reports that explain the temperature rise effect of the Er:YAG laser. Recently, the Er:YAG laser has been found to play an important role in temperature rises during the application on dentin. Methods: Caries-free primary mandibular molars were prepared to obtain dentin discs with 0.5, 1, 1.5, and 2 mm thicknesses (n = 10). These discs were placed between the Teflon mold cylinders of a temperature test apparatus. We preferred three pulse repetition rates of 10, 15, and 20 Hz with an energy density of 12.7 J/cm(2) and a 230 mu s pulse duration. All dentin discs were irradiated for 30 sec by the Er:YAG laser. Temperature rises were recorded using an L-type thermocouple and universal data loggers/scanners (E-680, Elimko Co., Turkey). Data were analyzed by two-way ANOVA and Tukey tests. Results: Whereas the lowest temperature rise (0.44 +/- 0.09 degrees C) was measured from a 10 Hz pulse repetition rate at a dentin thickness of 2 mm, the highest temperature rise (3.86 +/- 0.43 degrees C) was measured from a 20 Hz pulse repetition rate at a 0.5 mm dentin thickness. Conclusions: Temperature rise did not reach critical value for pulpal injury in any primary dentin thicknesses irradiated by a high repetition rate of the Er:YAG laser.

Açıklama

Anahtar Kelimeler

Kaynak

PHOTOMEDICINE AND LASER SURGERY

WoS Q Değeri

Q2

Scopus Q Değeri

N/A

Cilt

31

Sayı

5

Künye